Whakamātautau 0563 2051

Pukapuka Tohutohu mō te Mita pH Pae-T a te Testo 205

Tauira: 0563 2051

1. Kupu Whakataki

The Testo 205 is a robust handheld pH and temperature measuring instrument designed for use in semi-solid media. It is particularly suitable for applications in food production, such as monitoring the quality and maturation of meat in butcher shops, slaughterhouses, and the meat-processing industry. This manual provides essential information for the proper setup, operation, and maintenance of your Testo 205 device.

Testo 205 Handheld T-bar pH Meter

Figure 1: The Testo 205 Handheld T-bar pH Meter, showing its display and controls.

2. Nga waahanga matua

  • Simultaneous measurement of pH value and temperature in semi-solid media.
  • Integrated temperature sensor for automatic temperature compensation, ensuring accurate pH readings.
  • Maintenance-free pH probe with gel electrolyte, designed to be leak-proof and resistant to dirt.
  • Hole diaphragm on the pH probe for quick and reliable measurements.
  • User-replaceable combined pH measuring tip and temperature probe.
  • Two-line, illuminated LCD display showing both pH and temperature readings.
  • Automatic final value recognition (auto-hold function) with an audible advisory tone when readings stabilize.
  • Storage cap filled with gel electrolyte to maintain probe hydration and extend its lifespan.
  • Includes a belt/wall bracket for secure storage.
  • Supports one, two, or three-point calibration.

3. Tatūnga me te Whakamahi Tuatahi

3.1. Tāuta Pūhiko

The Testo 205 uses 4 x LR44 button cell batteries. To install or replace batteries:

  1. Kimihia te uhi o te wahanga pākahiko kei muri o te taputapu.
  2. Open the cover using a suitable tool (e.g., a small screwdriver or coin).
  3. Insert the 4 LR44 button cell batteries, ensuring correct polarity.
  4. Katia kia mau te uhi o te wahanga pākahiko.

3.2. Preparing the pH Probe

Before first use, or after prolonged storage, ensure the pH probe is properly conditioned.

  1. Remove the storage cap from the pH probe. The cap contains gel electrolyte to keep the probe hydrated.
  2. Rinse the probe tip with distilled water.
  3. Proceed with calibration as described in Section 4.1.

4. Nga Tohutohu Whakahaere

4.1. Whakatau

Regular calibration is crucial for accurate pH measurements. The Testo 205 supports one, two, or three-point calibration using buffer solutions.

  1. Pehia te KAUPAPA paatene ki te huri i te taputapu.
  2. Pehia te CAL button to enter calibration mode.
  3. Immerse the pH probe into the first calibration buffer solution (e.g., pH 7.00). Ensure the probe tip is fully submerged.
  4. The device will automatically detect the buffer and begin calibration. Wait for the reading to stabilize and the device to confirm calibration (e.g., by displaying "CAL OK" or a stable value).
  5. Horoia te tirotiro ki te wai matutu.
  6. For two or three-point calibration, repeat steps 3-5 with additional buffer solutions (e.g., pH 4.01 and pH 10.01).
  7. Exit calibration mode by pressing the CAL button again or waiting for automatic exit.

Tuhipoka: For economical use of calibration solution, Testo offers calibration dosing bottles with integrated dosing apertures.

4.2. Te Whanganga i tetahi Ine

To measure pH and temperature in semi-solid media:

  1. Ensure the device is calibrated and the probe is clean.
  2. Pehia te KAUPAPA paatene ki te huri i te taputapu.
  3. Carefully insert the pH probe into the semi-solid medium (e.g., meat). Ensure good contact and sufficient depth for accurate readings.
  4. The illuminated display will show both the pH value and temperature simultaneously.
  5. The automatic final value recognition (auto-hold) will capture the reading once it stabilizes. An advisory tone will sound to indicate stability.
  6. To manually hold a reading, press the KAUPAPA button during measurement. Press again to release.
  7. After measurement, rinse the probe thoroughly with distilled water.
Testo 205 pH Meter in use for meat measurement

Figure 2: The Testo 205 being used to measure the pH of a semi-solid food product, demonstrating its application.

4.3. Patene Aratau

Ko te MODE button allows you to switch between different display modes or access additional functions, if available. Refer to the on-screen prompts for specific options.

5. Tiaki

5.1. Probe Cleaning and Storage

Proper cleaning and storage are essential for maintaining the accuracy and lifespan of the pH probe.

  • I muri i ia whakamahinga: Rinse the pH probe thoroughly with distilled water to remove any residue from the measured medium.
  • Mo te rokiroki: Always place the storage cap, filled with gel electrolyte, back onto the pH probe. This prevents the probe from drying out and ensures its readiness for future measurements.
  • Te horoi hohonu: If the probe becomes heavily soiled, consult specialized cleaning solutions for pH electrodes. Avoid abrasive materials.

5.2. Replacing the pH Probe

The combined pH measuring tip with temperature probe is user-replaceable. If the probe's performance degrades significantly despite proper cleaning and calibration, it may need replacement.

  1. Gently twist and pull the old probe assembly from the main unit.
  2. Align the new probe assembly with the connector on the main unit.
  3. Push and twist gently until the new probe is securely attached.
  4. After replacing the probe, perform a full calibration as described in Section 4.1.

5.3. Whakakapinga Pūhiko

When the battery indicator appears on the display, replace the batteries as described in Section 3.1.

6. Te Raru

Raruraru Take pea Rongoā
Ko nga korero pH hē Probe not calibrated; dirty probe; dried-out probe; expired buffer solutions; damaged probe. Perform calibration (Section 4.1); clean probe (Section 5.1); rehydrate probe in storage solution; use fresh buffer solutions; replace probe (Section 5.2).
Te wa whakautu puhoi Dirty probe; dried-out probe; low battery. Clean probe (Section 5.1); rehydrate probe; replace batteries (Section 3.1).
Kaore te taputapu e ka ana Nga pākahiko mate; hē te tāutanga pūhiko. Whakakapia ngā pākahiko (Wāhanga 3.1); tirohia te polarity o te pākahiko.
Display shows "Err" or similar error message Probe issue; internal fault. Check probe connection; try cleaning/recalibrating; if problem persists, contact Testo support.

7. Whakatakotoranga Hangarau

Tawhā Uara
pH - Electrode
Awhe Ine 0 ki te 14 pH
Tika ±0.02 pH
Whakatau 0.01 pH
Temperature - NTC
Awhe Ine 0 to +60 °C (Short-term to +80 °C max. 5 min)
Tika ±0.4 °C
Whakatau 0.1 °C
Raraunga Hangarau Whānui
Paemahana Rokiroki -20 ki te +70 °C
Te Waahi Mahi 0 ki te +50 °C
Taumaha 135 g
Ahu (L x W x H) 145 x 38 x 167 mm
Momo Pūhiko 4 x Button cell LR44
Ora Pūhiko 80 h (Auto Off 10 Min)
Momo Whakaatu LCD, 2 lines

8. Pūtāhui me te Tautoko

8.1. Nga Korero Whakaputanga

Testo products are manufactured to high-quality standards. For specific warranty terms and conditions, please refer to the warranty card included with your product or visit the official Testo webpae. Puritia to tohu hoko mo nga kereme whakamana.

8.2. Tautoko Kiritaki

If you encounter any issues not covered in this manual or require further assistance, please contact Testo customer support. Contact details can typically be found on the Testo webte waahi kei roto ranei i te kohinga hua.

Mo etahi atu korero, tirohia te Testo Store on Amazon.

Tuhinga e pa ana - 0563 2051

Preview testo 風量ファンネルセット & フローストレーナー 取扱説明書 - 高精度な風鮇
whakamatautau 417, 400, 440シリーズ用風量ファンネルセットおよびフローストレーナーの取扱説明書。換気システムの給排気口での風量測定方法、接続、仕様についぬ。
Preview Testo 270 BT: Te Ara Tika ki te Ine - Aratohu Mahi me te Tiaki
Nga tohutohu tino nui mo te whakahaere me te pupuri i te whakamatautau hinu parai Testo 270 BT. Ako mo te whakarite, te ine tika, nga tikanga horoi, me te whakamatautau tika, tae atu ki te whakauru ki te testo Smart App.
Preview Testo 417 Vane Anemometer: Pukapuka Kaiwhakamahi me nga Raraunga Hangarau
He pukapuka whakamahi matawhānui mo te Testo 417 vane anemometer, e hipoki ana i te tikanga o te whakamahinga, te mahi, nga aratohu haumaru, nga whakaritenga hangarau, me nga taputapu. Akohia me pehea te ine tika i te tere rere me te pāmahana.
Preview Pukapuka Tohutohu mō te Whakamātautau Hinu Parai-Hohonu Testo 270 BT
Kei roto i tēnei pukapuka tohutohu ngā kōrero taipitopito mō te Testo 270 BT, he taputapu whakamātautau hinu parai hōhonu i hangaia hei ine tere me te tika i te kounga o te hinu tunu kai. Akohia ōna āhuatanga, te whakahaere, te tiaki, me te rapurongoā hei whakarite i te mahi tino pai me te haumaru kai.
Preview Testo Smart Probes Manual Kaiwhakamahi: Nga ahuatanga, nga korero, me te mahi
He pukapuka whakamahi matawhānui mo Testo Smart Probes, e whakaatu ana i nga ahuatanga, nga korero hangarau, nga aratohu haumaru, me nga tikanga whakahaere mo nga tauira penei i te testo 405i, 410i, 510i, 549i, 605i, 115i, 905i, me 805i. Kei roto i nga hua kua pauview, raraunga hangarau, me nga korero whakaaetanga.
Preview Testo Smart Probes Manual Kaiwhakamahi me Raraunga Hangarau
He aratohu matawhānui ki Testo Smart Probes, te whakamaarama i te tatūnga, te whakamahinga taupānga, nga tono ine, te rapurongoa, me nga whakaritenga hangarau mo nga tauira penei i te testo 115i, 405i, 510i, me etahi atu.