HACH 2100Q01

Hach 2100Q Portable Turbidimeter Instruction Manual

Tauira: 2100Q01

Kupu Whakataki

The Hach 2100Q Portable Turbidimeter is designed for accurate and reliable turbidity measurements in various environments, including field testing, plant monitoring, and water quality analysis for regulatory compliance. This manual provides essential information for the proper setup, operation, and maintenance of your 2100Q Turbidimeter.

Hach 2100Q Turbidimeter Kawe

Image: The Hach 2100Q Portable Turbidimeter, a compact device with a digital display showing a turbidity reading of 0.10 NTU.

Āhuatanga matua:

  • EPA Compliant Design: Meets USEPA Method 180.1 criteria for regulatory reporting.
  • Ine tika: Measures turbidity across a 0-1000 NTU range.
  • Guided Calibration: On-screen assisted calibration, including Rapidcal™ option.
  • Takitaki Raraunga: Stores up to 500 measurements with detailed information.
  • Kawe me te Roa: Rugged design suitable for field and plant operations.

Tatūnga

1. Te Wetewete me te Tirotiro

Carefully unpack all components from the carrying case. Verify that all items listed in the packing slip are present and undamaged. Report any missing or damaged items to Hach customer service immediately.

2. Tāuta Pūhiko

The Hach 2100Q requires four (4) AA batteries for operation.

  1. Locate the battery compartment cover on the back of the instrument.
  2. Open the cover by sliding or unlatching it according to the indicator.
  3. Whakauruhia nga pākahiko AA e wha, kia tika te polarity (+/-) kua tohua ki roto i te waahanga.
  4. Me tino kati te taupoki o te waahanga o te papahiko.

3. Hia Tuatahi

Press the power button (usually indicated by a circle with a vertical line) to turn on the turbidimeter. The display should illuminate, and the device will perform a self-check.

Nga Tohutohu Whakahaere

1. Whakatau

Regular calibration ensures accurate turbidity measurements. The 2100Q features on-screen guided calibration.

  1. Turn on the turbidimeter.
  2. Navigate to the "Calibrate" or "Calibration" menu option using the arrow keys.
  3. Follow the on-screen prompts to insert the required calibration standards (e.g., 20 NTU, 100 NTU, 800 NTU).
  4. For quick low-level calibration, utilize the Rapidcal™ option if available.
  5. Press "Read" or "Done" as prompted after each standard is inserted and measured.
  6. Ensure the calibration is successfully completed and saved.
Hach 2100Q Turbidimeter Calibration Screen

Image: Close-up of the Hach 2100Q screen during calibration, showing "Cal: StablCal® 99.4 NTU" and options for different NTU standards.

Inserting a sample vial into Hach 2100Q Turbidimeter

Image: A hand inserting a sample vial into the measurement cell of the Hach 2100Q Turbidimeter for calibration or measurement.

2. Te Whanganga i tetahi Ine

The 2100Q measures turbidity in Nephelometric Turbidity Units (NTU).

  1. Ensure the turbidimeter is calibrated.
  2. Whakapaia to sample according to standard laboratory practices, ensuring the sample cell is clean and free of air bubbles or scratches.
  3. Whakauruhia te sample cell into the measurement compartment, aligning any marks as instructed by the device or manual.
  4. Close the lid of the measurement compartment.
  5. Press the "Read" button. The device will display the turbidity reading in NTU.
Hach 2100Q Turbidimeter held in hand showing a reading

Image: A hand holding the Hach 2100Q Turbidimeter, displaying a turbidity reading of 103 NTU on its screen.

3. Data Logging and Transfer

The 2100Q automatically stores up to 500 measurements. Each entry includes a timestamp, operator ID, and calibration status.

  • ViewRaraunga: Access the "Data Log" or "History" menu option to review inenga rongoa.
  • Whakawhiti Raraunga: For transferring data to a computer, an optional USB+Power module is available. Connect the module to the turbidimeter and your computer, then follow the software instructions for data export.

Te tiaki

1. Te horoi i te taputapu

Keep the exterior of the turbidimeter clean using a soft, damp cloth. Do not use abrasive cleaners or solvents. Ensure the measurement cell compartment is free of dust and debris.

2. Sample Cell Care

Maama sample cells thoroughly before each use. Fingerprints, scratches, or residue on the cell can affect measurement accuracy. Use a lint-free cloth to wipe the exterior of the cell.

3. Whakakapinga Pūhiko

Replace batteries when the low battery indicator appears on the display. Refer to the "Battery Installation" section under Setup for instructions.

4. Optical System Care

The 2100Q uses a two-detector optical system. Avoid touching the optical components inside the measurement compartment. If cleaning is necessary, consult the full Hach 2100Q user manual for specific instructions to prevent damage.

Raparongoā

This section addresses common issues you might encounter with your Hach 2100Q Turbidimeter.

RaruraruTake peaRongoā
Kaore te taputapu e kaha ana.Nga pākahiko kua mate, kua he ranei te whakaurunga.Whakakapihia nga pākahiko, kia tika te polarity.
Ko nga panui kaore i te tika, kaore ranei i te rite.
  • Dirty or scratched sampte pūtau.
  • Ko te taputapu kaore i te whakatikatika.
  • Ngā mirumiru hau i roto i te sample.
  • Maama sample cell thoroughly.
  • Whakahaerehia he whakatikatika katoa.
  • Degas sample if necessary; tap cell gently to remove bubbles.
Ko te karere "Hapa" kei te whakaatu.Specific error code will indicate the issue.Refer to the comprehensive Hach 2100Q user manual for specific error code interpretations and solutions.

If you encounter issues not listed here or if the suggested solutions do not resolve the problem, please contact Hach Technical Support.

Whakatakotoranga

Below are the technical specifications for the Hach 2100Q Portable Turbidimeter (Model 2100Q01).

ĀhuahiraTaipitopito
Tau tauira2100Q01
Awhe Ine0 - 1000 NTU
TautukungaMeets USEPA Method 180.1 criteria
Puna MaamaTungsten Filament Lamp
Puna ManaE 4 nga papahiko AA e hiahiatia ana
Te Raraunga Raraunga RaukahaTae atu ki te 500 inenga
Pūnaha OpticalTwo-detector system for color and stray light compensation
Ahu (Ahua)16 x 10 x 8 inihi (Ngā Ahu Puka)
Taumaha (Ahua.)4 Pauna (Taimaha Puke)

Pūtāhui me te Tautoko

Nga korero Whakaputanga

Specific warranty terms and conditions for the Hach 2100Q Portable Turbidimeter are provided with the product documentation at the time of purchase. Please refer to the warranty card or the full user manual for detailed information regarding coverage, duration, and claims procedures. For the most current warranty details, contact Hach Company directly.

Tautoko Hangarau

For technical assistance, troubleshooting guidance, or service inquiries, please contact Hach Company customer support.

  • Kaihanga: HACH COMPANY
  • Webpae: www.hach.com (Please visit the official Hach website for contact information, support resources, and additional documentation.)

Rauemi Tāpiri

No official product videos from the seller were found in the provided data. For further operational guidance or visual demonstrations, please refer to the official Hach Company webpaetukutuku, whakapā atu ranei ki a raatau hongere tautoko.

Technician performing water quality testing with a portable device

Image: A technician in a hard hat and safety glasses using a portable water quality testing device in an industrial setting, demonstrating typical field use.

Tuhinga e pa ana - 2100Q01

Preview Pukapuka Kaiwhakamahi mō ngā Ine Turbidimita Kawe Hach 2100Q me te 2100Qis
Kei roto i tēnei pukapuka kaiwhakamahi ngā tohutohu whānui mō te whakahaere me te tiaki i ngā turbidimeter kawe Hach 2100Q me te 2100Qis. Akohia ngā whakatakotoranga, te tāutanga, te atanga kaiwhakamahi, te whakahaere matatau, me te rapurongoā.
Preview HACH TL2360 Basic User Manual - Turbidity Measurement Guide
User manual for the HACH TL2360 laboratory turbidimeter. Learn about installation, operation, calibration, and maintenance for accurate water turbidity measurements.
Preview Hach Model 2100N Laboratory Turbidimeter Instruction Manual
This instruction manual provides comprehensive guidance for the Hach Model 2100N Laboratory Turbidimeter, covering its operation, specifications, measurement procedures, calibration, maintenance, and troubleshooting for accurate turbidity analysis in laboratory environments.
Preview Hach Methods Quick Reference Guide: Chemical Analysis Instruments and Reagents
A comprehensive guide to Hach's chemical analysis methods, detailing test parameters, EPA approvals, instrument compatibility, and product numbers for water quality testing.
Preview HACH Conductivity Meter: USEPA Direct Measurement Method 8160 Guide
This guide details the HACH conductivity measurement method (USEPA 0.01 µS/cm to 200.00 mS/cm, Method 8160) for water and wastewater analysis. It covers instrument setup, test preparation, unit conversions, interferences, accuracy checks, and summary of the method.
Preview HACH FH950 Portable Flow Meter Basic User Manual
This document is the Basic User Manual for the HACH FH950 portable flow meter. It provides essential information for operating, installing, and maintaining the device for accurate water flow and depth measurements in various environments.