Elitech-logo

Elitech LogEt 6 Raraunga Raraunga Temperature

Elitech-LogEt-6-Temperature-Data-Logger-product

Nga tohutohu ahuru me nga whakatupatotanga

Nga tohutohu haumaru
To ensure that you install and use this product correctly, please read and strictly comply with the following terms:

Pūhiko

  • Please use original or technically compatible batteries. Do not use batteries that do not meet product specifications to prevent damage to equipment or other malfunctions.
  • Do not disassemble, squeeze, hit, or heat batteries privately, and do not place batteries in fire, as this may cause battery explosions and fires

Taonga mana o waho:

  • When external power supply is required, please use the power adapter configured for this device.
  • Do not use other power adapters that do not meet the external connection technical specifications, otherwise it may damage the instrument and equipment, and even cause a fire power supply.

Taputapu

  • In flammable and explosive gas environments, please use this equipment in accordance with compliance requirements; otherwise, it may cause explosions and fires
  • During the use of the device, if the device emits a burnt or other odor, the power supply should be immediately cut off and the manufacturer or supplier should be contacted in a timely manner

Te aro

  • If the equipment is not used for a long time, it should be removed and stored in the packaging box in a dry and cool environment
  • Users are not allowed to make any unauthorized changes to this device. Any unauthorized changes may affect the accuracy of the device and even damage it
  • Do not use the device in harsh environments to avoid instrument short circuits, burns, and other malfunctions caused by adverse weather conditions such as rain and lightning
  • When the recorder is offline (no data upload), please confirm the networking status of the device
  • Please use within the measurement range of the recorder
  • Please mistakenly place the external probe directly in a liquid environment to avoid equipment damage
  • Please do not forcefully impact this recorder
  • The measured values of the recorder may be affected by the following factors

Hapa pāmahana
Placed in the measurement environment for too short a stable time Close to heat sources, cold sources, or directly in a water soaked state

Humidity error

  • Placed in the measurement environment for too short a stable time
  • Long term exposure to steam, water mist, water curtains, or condensation environments

Parahanga:

  • Being in a dusty or other polluted environment

Ka mutu te huaview me te ahua

LogEt 6 is a disposable dry ice recorder with a built-in probe. Equipped with ultra-low temperature sensors as standard, the entire machine can directly record temperature in an environment of -85 ° C. It can export reports or automatically generate PDF files through data center software, with 16000 sets of storage, 30 days of battery life, 2 years of shelf life, and support for shadow data function before and after. Widely used in freezers, medical insulation boxes, laboratories, and other scenarios.Elitech-LogEt-6-Temperature-Data-Logger-fig-1

  1. Atanga USB
  2. LED
  3. mata LCD
  4. Patene Kati
  5. Tīmata pātene
  6. External probe interface (reserved)
  7. Pupuri pākahikoElitech-LogEt-6-Temperature-Data-Logger-fig-2
  1. max/min/avr/MTK/log quantity
  2. rohe runga
  3. rohe raro
  4. huru mahi
  5. taumata pākahiko
  6. Temperature/humidity values

Model list and specifications

Rarangi Tauira

Tauira Takiuru 6
whirihoranga PT
probe Type Built in ultra-low temperature (dry ice)
Main sensor parameters pūoko pāmahana
Awhe ine pāmahana -85℃~50℃
Te inenga tika ±0.5℃(-20℃~40℃);±1℃(other)
Taumira pāmahana 0.1℃

Nga whakakitenga

Te kaha rokiroki 16000
Atanga Raraunga USB A
Pūhiko LS14250 lithium battery
Papa-ora 2Tau
Te ora o te pākahiko 30Days(log interval5min)
Communication module/mode Atanga USB
Koeke tiaki IP65
Product dimensions (height*length*width) 100*46*19mm
Taumaha 60g
Te taiao mahi -85˚C~50˚C
Patene tinana Tīmata/Kati
Tikanga whakaoho LED
Adaptation software Elitechlog

Nga tohutohu whakahaere

Nga mahi matua
Start button: In normal display mode, press the start button to display the current temperature, and continue pressing to switch temperature data, date and time, number of records, maximum value, and minimum value; Press and hold for 5 seconds to start recording; Stop button: Long press for 5 seconds to stop recording;

Legend status indication

Elitech-LogEt-6-Temperature-Data-Logger-fig-3

Te mahi rorohiko

Tikiake pūmanawa
Download and install the “Jingchuang Data Center” software www.e-elitech.com/xiazaizhongxin (China), www.elitechlog.com/software (other countries)

Panui raraunga

  1. Connect the computer using a USB interface
  2. Open the “Jingchuang Data Center” and wait for the link to complete.
  3. Open the PDF file directly on my computer or open the summary/historical data of “Jingchuang Data Center” to view nga raraunga.

Notes
Data can be directly read without the need for the “Jingchuang Data Center” software After the device is taken out of a low-temperature environment, it needs to be left to stand for a period of time and wait for it to return to room temperature before the LCD screen displays data

Kaweake raraunga
Click on Summary/Historical Data, select Export Data, and choose the format to export

Rarangi taputapu

  • Host * 1, instruction manual (electronic version), calibration certificate (electronic version)

Rarangi me te ratonga muri-hoko

Pūtāhui hua: 1 tau
This product can be returned, replaced or repaired within 7 days from the date of purchase due to performance failure. The warranty period is one year from the date of purchase (based on valid purchase receipts). During the warranty period, any malfunctions caused by quality problems of the product under normal use will be repaired free of charge. The shipping cost for repairing the product shall be borne unilaterally by the sender.

If any of the following situations are met, it is not within the scope of the after-sales guarantee:

  1. Damage caused by force majeure, such as earthquakes, fires, floods, typhoons, etc;
  2. Damage caused by human factors, such as improper use, maintenance, and storage by consumers.
  3. Damage caused by maintenance, disassembly, etc. by non-precision service personnel or authorized service providers;
  4. Exceeding the warranty period and scope;
  5. Natural wear and tear, consumption, and aging of products (such as shells, cables, etc.).

FAQ

P: Me aha ahau mena ka puta te haunga o te taputapu?
A: Immediately disconnect the power supply and contact the manufacturer or supplier for assistance.

Q: Me pehea taku tiimata ki te tuhi i runga i te taputapu?
A: Press and hold the Start button for 5 seconds to initiate recording.

Tuhinga / Rauemi

Elitech LogEt 6 Raraunga Raraunga Temperature [pdf] Pukapuka Tohutohu
01, 02, LogEt 6 Raraunga Raraunga Paemahana, LogEt 6, Raraunga Raraunga Paemahana, Raraunga Raraunga, Raraunga Raraunga

Tohutoro

Waiho he korero